Appears in
IEEE Transactions on Intstrumentation and Measurement
Vol. 44, No. 2, April 1995
Abstract
The single ion Penning trap mass spectrometer at M.I.T. now compares masses with an accuracy of 0.1 part per billion. We have created a table of fundamental atomic masses and made measurements useful for calibrating the X-ray wavelength standard, and determining Avogadro's number, the molar Planck constant, and the fine structure constant.
Full paper as PDF, 188792 bytes.